| Manufacturer: | UWE | 
| Model(s): | PM | 
| Date updated: | 14.02.00 | 
1. Switch off.
2. Press and hold [MODE] key. Press [ON/ZERO] key.
3. Release [MODE] key.
4. Display shows [CAL?]
5. Press [MODE] key. The display shows [YE5], then [LoAd] followed by [XXXX] or [YYYY]. These are the values of the calibration weights to be used.
6. Place relevant calibration weight on plate.
7. After a pause the display will show [donE] and return to weighing mode.