CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: UWE
Model(s): OAP
Date updated: 04.09.00

1. Switch off.

2. Press and hold [NC/EURO] key. Press [ON/ZERO] key.

3. Release [NC/EURO] key.

4. Display shows [CAL?]

5. Press [NC/EURO] key. The display shows [YE5], then [LoAd] followed by [XXXX] or [YYYY]. These are the values of the calibration weights to be used.

6. Place relevant calibration weight on plate.

7. After a pause the display will show [donE] and return to weighing mode.