| Manufacturer: | UWE | 
| Model(s): | NKS | 
| Date updated: | 14.02.00 | 
SIMPLE CALIBRATION
1. Switch on scale.
2. Press [ZERO] key and while holding press [ON] key.
3. The display will show some figures. Press [ZERO] key.
4. Display will show zero count (NKS-2000 around 2500-4000, NKS-4000 around 1000-2500).
5. Press [ZERO] key to zero counts.
6. Place full capacity of scale on plate and press [ZERO] key again
7. Display will show "CAL", "8888", 2000. The calibration is complete.