CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: UWE
Model(s): NJW
Date updated: 14.02.00

1. Make sure machine is off.

2. Press [PCS/WT} key and while pressing ,switch on.

3. Don’t release [PCS/WT] key until display shows [CAL. ZRO].

4. Wait a few seconds. The display will show [CAL. SPN].

5. The display will then show two figures representing the choice of calibration weights.

6. Place either one of the weights (preferably the higher one) on the scale.

7. The scale will automatically calibrate and display [DONE] when complete.

8. The display will then count down in numerical sequence and revert to weighing mode.