CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: UWE
Model(s): HS
Date updated: 17.02.00

1. Switch off.

2. Press and hold [HOLD] key. Press [ON/ZERO] key.

3. Release [HOLD] key. Display shows [CAL?]

4. Press [HOLD] key. The display shows [YE5], then [LoAd] followed by [XXXX] or [YYYY]. These are the values of the calibration weights to be used.

5. Hang relevant calibration weight on hook or place in pan (if supplied). After a pause the display will show [donE] and return to weighing mode.