Manufacturer: | UWE |
Model(s): | HS |
Date updated: | 17.02.00 |
1. Switch off.
2. Press and hold [HOLD] key. Press [ON/ZERO] key.
3. Release [HOLD] key. Display shows [CAL?]
4. Press [HOLD] key. The display shows [YE5], then [LoAd] followed by [XXXX] or [YYYY]. These are the values of the calibration weights to be used.
5. Hang relevant calibration weight on hook or place in pan (if supplied). After a pause the display will show [donE] and return to weighing mode.