CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: UWE
Model(s): 1707 Indicator
Date updated: 17.02.00

1. Switch off.

2. Press & hold [SET] key. Press [ON/ZERO] key.

3. Display will show [CAL-0 kg] to show scale zero is being calibrated.

4. After a pause the display will show [00000 kg] with one "0" flashing. Set the span calibration weight by using the[MODE] key to move the flashing "0" from left to right. Use the [TARE] key to increase the figure and the [ON/ZERO] key to decrease the figure.

5. Place the chosen calibration weight on the platform.

6. After a pause the display will show [donE] and calibration is complete