CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: TANITA
Model(s): TLD-612
Date updated: 11.02.00

1. While pressing the [ ® T ] key, press the [ I / ® 0 ] to turn the power on.

2. Display should show [ 88888 ] followed by a number [ X ].

3. Press the [ TARGET ] key. Display should show [ 0AA ] followed by [ 1AA ].

4. Press the [ TARGET ] key again. Display should show [ 2AA ].

5. Place 1kg weight on the weight plate.

6. After allowing the machine to settle press the [ TARGET ] key. Display should show [ ----- ].

7. Calibration complete. Switch off and on again & re-check with test weight.