Manufacturer: | TANITA |
Model(s): | TLD-612 |
Date updated: | 11.02.00 |
1. While pressing the [ ® T ] key, press the [ I / ® 0 ] to turn the power on.
2. Display should show [ 88888 ] followed by a number [ X ].
3. Press the [ TARGET ] key. Display should show [ 0AA ] followed by [ 1AA ].
4. Press the [ TARGET ] key again. Display should show [ 2AA ].
5. Place 1kg weight on the weight plate.
6. After allowing the machine to settle press the [ TARGET ] key. Display should show [ ----- ].
7. Calibration complete. Switch off and on again & re-check with test weight.