CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: Shinko
Model(s): PF
Date updated: 18.02.00

1. Press [F] key until display shows [Func] followed by [CAL].

2. Press and hold [T] key, press [F] key and release both together.

3. The display shows [unit X] where "X" is the weighing unit previously set into the scale as shown in the following table.

Unit 1 = "mg"

Unit 2 = "g" Unit 8= "dwt"

Unit 3 = "kg" Unit 9 = "gr"

Unit 4 = "ct" Unit A = "tl" (Hong Kong)

Unit 5 = "oz" Unit b = "tl" (Singapore, Malaysia)

Unit 6 = "lb" Unit c = "tl" (Taiwan)

Unit 7 = "ozt" Unit d = "mom"

4. To change from one unit to another press[T] key. When unit is satisfactory press [F] key to store it.

5. The display should show [on 0 ]. Make sure nothing is on the pan.

6. The display will automatically advance to [ on F.S ]. Apply a reference weight which must be at least 50% of the capacity of the scale in the centre of the pan and the span will be adjusted automatically.

7. When calibration is complete the display will return to weighing mode.