Manufacturer: | Shinko |
Model(s): | DJ-E |
Date updated: | 18.02.00 |
1. Press [F] key until display shows [Func] followed by [CAL].
2. Press and hold [T] key, press [F] key and release both together.
3. The display starts blinking [on 0]. Make sure nothing is on the pan.
4. The display will automatically advance to [ on F.S ]. Apply a reference weight which must be at least 50% of the capacity of the scale in the centre of the pan, the display will start blinking and the span will be adjusted automatically.
5. When calibration is complete the display will return to weighing mode.