CALIBRATION INSTRUCTIONS

Index of Calibration Instructions | Stringer Home Page


Manufacturer: Shinko
Model(s): CGX
Date updated: 18.02.00

1. Press [MODE] key until display shows [Func] followed by [CAL].

2. Press and hold [T] key, press [MODE] key and release both together.

3. The display shows [unit X] where "X" is the weighing unit previously set into the scale as shown in the following table.

Unit 1 = "g"

Unit 2 = "kg"

Unit 3 = "ct"

Unit 4 = "oz"

Unit 5 = "lb"

4. To change from one unit to another press[T] key. When unit is satisfactory press [MODE] key to store it.

5. The display should show [on 0 ]. Make sure nothing is on the pan.

6. The display will automatically advance to [ on F.S ]. Apply the reference weight which must be at least 50% of the scale capacity in the centre of the pan and the span will be adjusted automatically.

7. When calibration is complete the display will return to weighing mode.