Manufacturer: | Shinko |
Model(s): | CGX |
Date updated: | 18.02.00 |
1. Press [MODE] key until display shows [Func] followed by [CAL].
2. Press and hold [T] key, press [MODE] key and release both together.
3. The display shows [unit X] where "X" is the weighing unit previously set into the scale as shown in the following table.
Unit 1 = "g"
Unit 2 = "kg"
Unit 3 = "ct"
Unit 4 = "oz"
Unit 5 = "lb"
4. To change from one unit to another press[T] key. When unit is satisfactory press [MODE] key to store it.
5. The display should show [on 0 ]. Make sure nothing is on the pan.
6. The display will automatically advance to [ on F.S ]. Apply the reference weight which must be at least 50% of the scale capacity in the centre of the pan and the span will be adjusted automatically.
7. When calibration is complete the display will return to weighing mode.