Manufacturer: | Shinko |
Model(s): | CG |
Date updated: | 18.02.00 |
1. Press [F] key until display shows [Func] followed by [CAL].
2. Press and hold [T] key, press [F] key and release both together.
3. The display shows [unit X] where "X" is the weighing unit previously set into the scale as shown in the following table.
Unit 1 = "g" Unit 7 = "dwt"
Unit 2 = "kg" Unit 8 = "gr"
Unit 3 = "ct" Unit 9 = "tl" (Hong Kong)
Unit 4 = "oz" Unit A = "tl" (Singapore, Malaysia)
Unit 5 = "lb" Unit b = "tl" (Taiwan)
Unit 6 = "ozt" Unit c = "mon"
4. To change from one unit to another press[T] key. When unit is satisfactory press [F] key to store it.
5. The display should show [on 0 ]. Make sure nothing is on the pan.
6. The display will automatically advance to [ on F.S ]. Apply a reference weight which must be greater than 50% of the capacity of the scale in the centre of the pan and the span will be adjusted automatically.
7. When calibration is complete the display will return to weighing mode.